Product Details
- 1 button testing - No setup required, just plug right in! Or, make a plug for your cordless product, and plug right in! EASY!
- Easy to identify PASS / visible and audible FAIL indication
- Auto Test Time (1 sec. / 60 sec.)+ - No need to set a timer or stop watch
- Safety Protected Start Switch -- Protects user from shock!
- Performs the Low Current Ground Continuity test at the same time
- Meets UL, CSA, CE, EN, IEC, MIL and all other international safety agency requirements as well as UL 120Kohm requirements
- Includes ISO/IEC 17025 Accredited Calibration Certificate, accredited by ACLASS (Internationally accepted results - equivalent to A2LA)
TECH NOTE: We get numerous calls from users who are looking for AC hipot testers with high output current capabilities. In almost every case, the user actually needed a DC hipot tester. When hipot testing products that incorporate EMI filters (for EMC regulations),
you need to use DC, not AC!
With AC, the Y capacitors in the EMI filter will conduct current, causing false failure indications and possibly damaging these capacitors. Most "high-tech" standards anticipate this and permit DC testing at 1.414 times the AC test voltage. This includes
IEC950, IEC1010, and IEC601based Standards.
Meets Requirements for Testing Standard(s) including but not limited to:
- IEC 60335
- IEC 60601
- IEC 60950
- IEC 61010
- UL 1703