Educated Design & Development has been named an company.
Top

LT-30HC Digital Leakage Current Tester LT-30HC Digital Leakage Current Tester






Product Details
  • Features product ground switching (grounding conductor maintained to device under test when not measuring leakage).
  • Automatic front receptacle disconnect when configured in the 240V test mode (prevents inadvertent connection of 120V when configured for 240V operation).
  • These are the only leakage current testers designed and supported by product safety experts.
  • Full technical support available by experienced safety engineers.
Input / Output Ratings 100 -120 / 200-240 VAC 30 A 50-60 Hz Dimensions
Weight
13.25 in. W x 11.5 in. D x 5.5 in. H
13 lbs.
Testing Tolerance 90 -132 / 180 - 264 VAC Measurement Ranges 2 mA/Resolution: 1.0 μA, Accuracy: ± 1% full scale
20 mA/Resolution: 10.0 μA, Accuracy: ±1% full scale
Leakage Current Measurement Switch Can be measured through:
  • product power cord ground conductor; Surface-to-Line 1/Line 2
  • OR Ground; Surface-to-Surface
Body Impedance Circuits 1500 Ω, 1% resistance shunted by 0.15 μF, 1% capacitance, compliant with standards requiring that body circuit
Display 3.5 in. digital LCD display