LT-601 and LT-601HC Digital Leakage Current Tester LT-601 and LT-601HC Digital Leakage Current Tester






Product Details

The LT-601 and LT-601HC meets UL, CSA, CE, EN, IEC, MIL standards... These devices also meet standards for Medical product safety testing!

A front panel switch permits Leakage Current to be measured:
  • Through product's power cord ground conductor
  • Surface/Patient-to-Ground
  • Patient-to-Patient
Input Ratings LT-601: 100-120/200-240 VAC 15 A 50-60 Hz
LT-601HC: 100-120/200-240 VAC 30 A 50-60 Hz
Testing tolerance: 90-132 / 180-264 VAC
Mechanical Size: 13.25 in. W x 11.5 in. D x 5.5 in. H
LT-601: Weight: 9 lbs.
LT-601HC Weight: 13 lbs.
Output Ratings Same as input ratings Measurement Ranges 200 μA/Resolution: 1.0 μA, Accuracy: ± 1% full scale
2000 μA/Resolution: 10.0 μA, Accuracy: ± 1% full scale
Display 3.5 in. digital LCD display Body Impedance Circuits Network in accordance with IEC 60601, EN 60601, CSA 601, UL 2601